Autoclaves > Top line verticals TLV-DUAL Series

TLV-DUAL Series

Vertical autoclaves with super-drying and fast cooling system

The TLV-DUAL Series vertical laboratory autoclaves combine our fast cooling and super-drying systems for faster sterilization of both solid and liquid loads.

  • Sterilization with final drying and fast cooling for solid loads and objects with complex geometries
  • Modern design with 7'' capacitive touchscreen display.
  • Advanced cloud-based connectivity that complies with FDA, GMP, and GLP standards

  • TLV-DUAL autoclave RAYPA 1%
  • TLV-DUAL autoclave RAYPA 2%
  • TLV-DUAL autoclave RAYPA 3%
  • TLV-DUAL autoclave RAYPA 5%
  • TLV-DUAL autoclave RAYPA 6%
  • TLV-DUAL autoclave RAYPA 7%
  • TLV-DUAL autoclave RAYPA 8%
  • TLV-DUAL autoclave RAYPA 9%
  • TLV-DUAL autoclave RAYPA 10%
  • TLV-DUAL autoclave RAYPA 11%
  • TLV-DUAL autoclave RAYPA 12%
  • TLV-DUAL autoclave RAYPA 13%

Vertical laboratory autoclaves with super-drying and fast cooling system

The TLV-DUAL Series autoclaves feature a super-drying system including a heating jacket and a high-performance vacuum pump, as well as a fast cooling system comprising a radial fan, water cooling coil, and air compressor. They are designed to provide advanced sterilization of solids and liquids in industrial and research settings within the pharmaceutical, chemical, and biotechnological industries, sterilizing liquids, culture media, hermetically sealed items, pre-filled syringes, glassware, porous solids, and wrapped objects. Top line vertical autoclaves enable fully digital operation. All models are equipped with the most advanced connectivity on the market and comply with the latest advances in electronic records and data control to be able to work under FDA, GMP and GLP environments.

RAYPA's TLV-DUAL Autoclave with super-drying and fast cooling system

Discover the most advanced technology for the sterilization of porous solids and bioreactors

Thanks to its super-drying system, which combines a vacuum pump and a heating jacket, you can sterilize porous solid loads, textiles and complex geometries more efficiently and quickly.

State-of-the-art technology for advanced liquid sterilization

Thanks to its fast cooling system consisting of a radial fan, water coil and air compressor, you will be able to sterilize liquid loads more quickly. Install up to 2 flexible probes and conduct F₀-controlled sterilizations.

TLV-DUAL autoclave RAYPA detail 2
TLV-DUAL autoclave RAYPA detail 3

Cloud-based connectivity and advanced technical support

Monitor sterilization cycles from anywhere. Schedule a service meeting from the controller itself and share a screen via TeamViewer®. Online diagnostics of equipment status.

Operation

Temperature

Pressure

TLVDUAL

A
B
C
D
E
TLV DUAL A
TLV DUAL B
TLV DUAL C
TLV DUAL D
TLV DUAL E
A. Prevacuum phase
In this phase the cold air in the chamber is mechanically purged at high speed by means of multiple vacuum pulses. At the same time, the steam generator is activated, which will inject saturated steam into the chamber. This process removes non-condensable gases from inside the chamber, eradicating the presence of cold air pockets and facilitating steam access to any geometry of the load.
B. Heating phase
In this phase, the steam generator continues to inject saturated steam into the chamber until the sterilization temperature is reached.
C. Sterilization phase
Upon reaching the sterilization temperature in the medium, the sterilization phase begins and the temperature is accurately sustained for the predefined duration.
D. Vacuum drying phase
In programs for solid loads, a fast vacuum drying phase begins using the vacuum pump and heating jacket to completely dry the load.
E. Cooling phase
Finally, a fast cooling phase begins, featuring pressure support along with the activation of the water cooling coils and internal radial fan to quickly cool the load. Optionally, the agar mode can be configured to maintain a constant temperature.

Operation

In this phase the cold air in the chamber is mechanically purged at high speed by means of multiple vacuum pulses. At the same time, the steam generator is activated, which will inject saturated steam into the chamber. This process removes non-condensable gases from inside the chamber, eradicating the presence of cold air pockets and facilitating steam access to any geometry of the load.

Temperature

Pressure

Pre-vacuum phase in RAYPA's TLV-DUAL autoclave%

In this phase, the steam generator continues to inject saturated steam into the chamber until the sterilization temperature is reached.

Temperature

Pressure

Heating phase in RAYPA's TLV-DUAL autoclave%

Upon reaching the sterilization temperature in the medium, the sterilization phase begins and the temperature is accurately sustained for the predefined duration.

Temperature

Pressure

RAYPA's TLV-DUAL autoclave sterilization step%

In programs for solid loads, a fast vacuum drying phase begins using the vacuum pump and heating jacket to completely dry the load.

Temperature

Pressure

Fase de %

Finally, a fast cooling phase begins, featuring pressure support along with the activation of the water cooling coils and internal radial fan to quickly cool the load. Optionally, the agar mode can be configured to maintain a constant temperature.

Temperature

Pressure

RAYPA's TLV-DUAL autoclave cooling phase%

Applications

Technical data

References
Total/usable volume of the chamber (L)
Usable dimensions of the chamber (Ø x H mm)
External dimensions (L x D x H mm)
Loading height (mm)
Net weight (Kg)
Power (W)
Voltage* (V)
Frequency (Hz)
*Other voltages and electrical configurations available on request.